My status
Manufacturer
Product search
April  25 , 2024  Thursday
Home> analysis meter >> analytical instrument >> ZD-2Automatic Titrator
  • Introduction
  • Specifications
ZD-2 Automatic Titrator, used chemical capacity analysis, also can be used with PH meter or High impedance mV meter
Use touch screen control, Chinese operation interface, simple using
Use module design, made up of titration device, control device and measuring device, achieve potential, titration, conductometric titration, and dead-stop titration
1、  measuring range
a) pH : (0.00~14.00)pH
b) mV : (0~±1999.0)mV
c) temperature :(-5.0~105.0)℃
2、error of electronic unit
a、PH:±0.01PH ± 1 bytes
b、mV:±0.03%FS± 1 bytes
c、temperature:±0.3℃± 1 bytes
3、input current of electronic unit: < 1×10-12A
4、input resistance of electronic unit: >3×1012Ω
5、stability of electronic unit: ±0.3 mV± 1 bytes/3h
6、sensitivity of control titration : ±2mV
7、error of burette capacity:    10ml burette:±0.025ml   20ml burette:±0.035ml
8、burette infusion or infusion rate: 55±10s
9、repeatability of titration analyzing : 0.2%
,