My status
Manufacturer
Product search
April  28 , 2024  Sunday
Home> analysis meter >> analytical instrument >> DWG-8025Asodium monitor
  • Introduction
  • Specifications
Used in power plant condensate, feed water, demineralized for continuous detection
Backlight display, Chinese menu, same time show sodium concentration and temperature
With (4~20)mA DC isolated output signal, output upper and lower limit of sodium concentration, free setting of upper, lower limit alarm. Contact volume 2A,220V AC
With RS232 or RS485 output interface, achieve digital output
Measuring material: plexiglass
1.measuring range: sodium concentration:(0.1~1000)μg/l ,temperature:(-10~120)°C
2.basic error: sodium concentration:±10%(reading)± 0.1μg/l,temperature: ±0.5°C ±1 bytes (0.0°C≤T≤60.0°C);±1.0°C ±bytes
3.stability of electronic unit: ±3%(reading)/24h
4.response time: <6min
5.temperature of measured solution: 5~45)°C
6.flow of measured water: 100ml/min
7.pressure of measured water: 100ml/min
8.entrance connection of measured water: external diameterΦ6×1 stainless steel or soft tube
exit connection of measured water: external diameterΦ10×1 stainless steel or soft tube
9.dimension, mm:electronic unit:144×144×150; measuring unit:628×228×224
10.hole size: mm:electronic unit:138×138;measuring unit:262×602,Weight: electronic unit: 2kg,measuring unit: 10kg
,