Sensitivity of TCD: S≥1000mV.ml/mg
Flame ionization detector(FID):Dt≤1×10-10g/s
Thermostatic chamber:
Temperature range of column box: 30℃-300℃
Temperature accuracy: ±0.1℃-0.2℃
Temperature control range of detector, vaporizer, conversion room : 50℃-399℃
Temperature error of active area: ≤2℃
Detector stability:
Baseline noise: ≤0.05mV
Baseline drift: ≤0.15mV/h
,