Thermal conductivity detector:> 2500mv ml/mg(nc16)
Baseline drift: ≦20μv/1Smin
Hydrogen flame detector: Mt<1×10-11g/s(nc16)
Baseline drift: ≦1×10-13A15min
Electron capture detector: Mt≦2×10-13g/s(rBHC)
Baseline drift: ≦1SHZ/15 min
Flame photometric detector: Mt(P)≦2×10-11 g/sec(methyl-parathion)
Baseline drift: ≦20μv/15min
Nitrogen and phosphorus detector: Mt(N)≦5×10-10 g/s(Azobenzene)
Mt(P)≦5×10-12 g/s
,