Configuration features With PROFIBUS-PA communication protocols, through the PC operating software (such as CommuwinII) for configuration
Input signal type Thermal resistance, all kinds of thermocouples, resistance, millivolt signal
range Depends on the sensor connection mode and the input signal
Output signal Physical data transmission (physical layer)
bus interface complies with the IEC61158-2
Default signal Status information in line with PROFIBUS-PA V3.0 Description
Filter Class 1 Digital Filter:0~60s
Current Compensation 10mA±1mA
Fault current 0mA
Data Transfer Rate 31.25KBit/sVoltage-mode
Signal code Manchester II
Response time 1s
Isolated voltage 2kVAC (input/output)
Power voltage 9~30VDCNon-Ex area,Polarity Protection; 9~15VDC Ex area,Polarity Protection
Long-term stability ≤0.1℃/year,or 0.05%/year
Protection level IP66(TMT184FA) IP65(TMT184FP)
Ambient temperature -40~+85℃