Explosion-proof mark ExdIIBT4
Qualified number GYB05113
Use area 1area、2 area、IIA、IIB,T1~T4 dangerous area
Configuration features With PROFIBUS-PA communication protocols, through the PC operating software (such as CommuwinII) for configuration
Input Signal Type Thermal resistance, all kinds of thermocouples, resistance, millivolt signal
range Depends on the sensor connection mode and the input signal
Conversion Accuracy ≤0.08%
Output signal analog 4~20mA 20~4mA
Fault signal status information in line with PROFIBUS-PA V3.0 Description
Current compensation 10mA ± 1mA
Fault current 0mA
Data transfer rate 31.25kBit / s voltage mode
Signal Code ManchesterII
Response time 1s
Display digit four digits
Isolated voltage 2kVAC (input/output)
Power voltage 9~30VDC
Long-term stability ≤0.1℃/year,or 0.05%/year
Protection level IP66
Ambient temperature -20℃~+40℃