Explosion-proof mark:Ex(ia)ⅡC
Transfer Accuracy:±0.2%×F.S
allowable input signal in dangerous are
thermocouple:K、E、S、B、J、T、R、N
millivoltage:0~20mV、0~75mV、0~200mV
output signal to safety area:
DC current:0~10mA、4~20mA、0~20mA
DC voltage:0~5V,1~5V,0~10V
output loading:DC:0~10mA≥1KΩ;4~20mA≥500Ω
NEPSI certification parameter(between 9 and 10 terminals):
Uo=8.7V,Io=27mA,Co=3.0μF,Lo=20mH
apply range:thermocouple K、E、S、B、J、T、R、N and millivotage sensor。
Cold Junction Compensation Accuracy:1℃
communication(safety area):RS232 or RS485 interface
power supply:DC 24V±10%
power consumption:1.2W(single-loop 20mA output);1.8W(dual-loop 20mA output )
case size: 22.5×100×115 mm